2012年3月8日星期四

The maintenance ,alarm and processing of blood analyzer

   Automatic Blood Coagulation Analyzer is purchased  from 2005 in our hospital.It has brought convenient to the work and provides the more accurate results to the clinical. But in use process all can't avoid for various reasons, have this or that kind of failure, but the failure rate just in different instruments. To guarantee the normal work of the instrument, to the instrument for regular maintenance is very important. Therefore, in order to put the instrument in the best state work, except the operating personnel to this instrument performance has a more comprehensive understanding of the staff also often to the instrument for some daily maintenance , so as to reduce the instrument in use process because of failure and affect the accuracy of the results, and reducing the instrument malfunction and produce the waste of resources. At the same time in during the work that the instrument has alarm to make timely corresponding processing. Now the methods of  the daily work, some simple maintenance, maintenance of automatic blood coagulation analyzer are concluded and peered learn together.

   1 Instrument Maintain   For keeping the ideal state of work, must be carefully safeguard. Maintenance is divided into daily, weekly, monthly, and every two weeks a year. The daily work is often do maintenance summarized as follows:


  1,1 Daily Maintenance    In order to get the best results, this machine has been in the state to boot. Reduce instrument "on" and "off" times. Before working every day to check  the amount of liquid whether enough, and reagent liquid surface height should ensure that no less than 210 cm high. Check waste liquid barrels of liquid surface height, confirmed that waste liquid to smoothly to waste in the barrel and, when necessary, to empty the waste liquid. To avoid causing waste liquid pipe of the jam by blood clot formation, every time to complete daily work to implement a waste liquid pipe cleaning process.


  1.2 Weekly Maintenance    Brush the instrument appearance, automatic vacuum device, sample than color dish interval part (not including than color dish holder) with  0.1 mol/L HCl and then use deionized water rinse. Clean needles to ensure that remove the needle protein and other internal and external sediments. When the specimens spatter to automatic vacuum device, or samples than color dish interval parts, to automatic vacuum sample device, color sensors and measuring cup than two of the room light path parts for cleaning. With 0.1 mol/L HCl dilute solution of clean cloth or after soaking wipe a swab than two color cup sensor vertical plane, then use deionized water cleaning and clean cloth or a swab dry.
Instruments in the daily work of the common fault and its corresponding processing methods and the cause of the problem is very complex.Not only is the instrument itself, also involves the instrument and apparatus operator in the external environment. As long as the operator, the instrument itself and the environment in the three any problems, it can lead to medical test instruments of failure. Therefore, the operator, the instrument, the environment they form a medical test instruments of the three important aspects. Now the operating automatic blood coagulation analyzer to some common warning and when its corresponding processing method summarized as follows:


2.1 The instrument shows "INCUBATION TEMP OUT OF RANGE"means the temperature of the color storehouse is high  (39 ~ 50 ℃) or low (20 ~ 36 ℃), correct temperature is 38.5 ℃. Treatment: if the instrument shows the temperature too high, then check the ventilators and around of exhaust equipment whether in the range of the instrument .If display temperature too low, then check the heater device and the temperature around, best is greater than 15 ℃, if correct mistakes after still exist, then contact please engineer.


2.2 The instrument shows "CYCLE ABORTED NO FLUSH OPTIC ERROR) means liquid finished, and light path ERROR occurred. The cases above except's participation,there is another solution,staff take the washing liquid into than liquid, the device also will alarm this ERROR.Treatment:When change the liquid,take care the right liquid.


2.3 The instrument shows "NOT IDENTIFIED SAMPLE IN POS" means samples and sample plate input the actual Numbers do not tally. Treatment: to check input samples and make the number of the actual equal to the input amount.

没有评论:

发表评论